UES240E-SPCZ/UES240F-SPCZ
- Compliant with medical and telecommunications security certifications
- 2 MOPP isolation protection
- Leakage current ≤100μA
- Meets Level 7 efficacy
- Power consumption ≤ 0.15W
- 5V/9V/12V/15V/20V/28V/36V/48V输出,最高240W
- Working altitude 5000 meters
- Meets USB PD3.2, AVS, PPS, and QC3.0 fast charging protocols.
- USB port/fixed cable design
Product Features
| Input voltage | 80-264 volts AC | Working temp. | -10 ~ 40℃ |
| Input Current | 4.0A, under 80VAC input condition | Storage temperature | -20 ~ 80℃ |
| Altitude | 5000 meter | (I.e. | 5~90% RH, non-condensing |
| Standby power consumption | ≤0.15W | mean time between failures | At an ambient temperature of 25°C, the mean time between failures (MTBF) under maximum load is 10 hours (according to SR-332 standard). |
| weight | 650±15 grams | Mechanical specifications | 150(L) 73(W) 32.5(H) mm |
| Application places | Suitable for medical, information and information equipment power adapters | Safety standards | CB(IEC60601-1/IEC60601-1-11/IEC62368-1)、TuV-SUD/Mark(EN60601-1/EN60601-1-11)、UL(ANSI/AAMI ES60601-1/60601-1-11/CAN/CSA C22.0 NO.60601-1/UL62368-1)、RCM(AS/NZS 62368)、CE(EN62368)、CCC(GB4943.1)、FCC(FCC PART 15)、KC(K62368)、PSE(J62368) |
| model | Voltage(V) | Current (A) | Power (W) | Ripple (maximum) mVpk-pk | Voltage tolerance | Linear and Negative Adjustment Rates | Average efficiency | Start delay time |
UES240E-SPCZ UES240F-SPCZ
| 5.0 | 0.01-3.00 | 15 | 250 | ± 5% | Linear regulation: ±2% Load regulation: ±5%
| 82.0 % | ≤3 seconds |
| 9.0 | 0.01-3.00 | 27 | 250 | ± 5% | 87.3 % | ≤3 seconds | ||
| 12.0 | 0.01-3.00 | 36 | 250 | ± 5% | 88.3 % | ≤3 seconds | ||
| 15.0 | 0.01-3.00 | 45 | 250 | ± 5% | 88.9 % | ≤3 seconds | ||
| 20.0 | 0.01-5.00 | 100 | 300 | ± 5% | 89.5 % | ≤3 seconds | ||
| 28.0 | 0.01-5.00 | 140 | 350 | ± 5% | 90.0 % | ≤3 seconds | ||
| 36.0 | 0.01-5.00 | 180 | 350 | ± 5% | 91.0 % | ≤3 seconds | ||
| 48.0 | 0.01-5.00 | 240 | 400 | ± 5% | 91.5 % | ≤3 seconds |
| Input voltage | 80-264 volts AC | Working temp. | -10 ~ 40℃ |
| Input Current | 4.0A, under 80VAC input condition | Storage temperature | -20 ~ 80℃ |
| Altitude | 5000 meter | (I.e. | 5~90% RH, non-condensing |
| Standby power consumption | ≤0.15W | mean time between failures | At an ambient temperature of 25°C, the mean time between failures (MTBF) under maximum load is 10 hours (according to SR-332 standard). |
| weight | 650±15 grams | Mechanical specifications | 150(L) 73(W) 32.5(H) mm |
| Application places | Suitable for medical, information and information equipment power adapters | Safety standards | CB(IEC60601-1/IEC60601-1-11/IEC62368-1)、TuV-SUD/Mark(EN60601-1/EN60601-1-11)、UL(ANSI/AAMI ES60601-1/60601-1-11/CAN/CSA C22.0 NO.60601-1/UL62368-1)、RCM(AS/NZS 62368)、CE(EN62368)、CCC(GB4943.1)、FCC(FCC PART 15)、KC(K62368)、PSE(J62368) |
| model | Voltage(V) | Current (A) | Power (W) | Ripple (maximum) mVpk-pk | Voltage tolerance | Linear and Negative Adjustment Rates | Average efficiency | Start delay time |
UES240E-SPCZ UES240F-SPCZ
| 5.0 | 0.01-3.00 | 15 | 250 | ± 5% | Linear regulation: ±2% Load regulation: ±5%
| 82.0 % | ≤3 seconds |
| 9.0 | 0.01-3.00 | 27 | 250 | ± 5% | 87.3 % | ≤3 seconds | ||
| 12.0 | 0.01-3.00 | 36 | 250 | ± 5% | 88.3 % | ≤3 seconds | ||
| 15.0 | 0.01-3.00 | 45 | 250 | ± 5% | 88.9 % | ≤3 seconds | ||
| 20.0 | 0.01-5.00 | 100 | 300 | ± 5% | 89.5 % | ≤3 seconds | ||
| 28.0 | 0.01-5.00 | 140 | 350 | ± 5% | 90.0 % | ≤3 seconds | ||
| 36.0 | 0.01-5.00 | 180 | 350 | ± 5% | 91.0 % | ≤3 seconds | ||
| 48.0 | 0.01-5.00 | 240 | 400 | ± 5% | 91.5 % | ≤3 seconds |
Product information
| Safety | CB, CE, cULus, FCC, TUV-Mark, UL |
|---|
Compare
Related Products
model
Output Voltage
Output Current
power
Application industry
Data sheet