UES310D-SPAZ

  • Compliant with communications and medical safety certifications
  • 2 MOPP isolation protection
  • Contact current ≤ 100
  • Meets Level 7 energy efficiency
  • Standby power ≤ 0.15W
  • 12VTo54VOutputUp to 310W
  • Working altitude 5000 meters
  • There are 5 AC socket types for input
  • GaN technology
  • Meets home medical standards
Product Features
Input voltage90-264VACWorking temp.0〜40 ℃        
Input Current5.0AStorage temperature-20~60℃       
Output voltage range± 5%Working humidity10~90%RH, no condensation       
Storage humidity5~90% RHmean time between failures > 100,000 hours Telcordia_SR-332, at 25°C ambient temperature.        
weightAbout 800gMechanical specifications73.5 (L)  42.5(W)  30.7(Hmm       
Application placesSuitable for medical and ITE equipment power adaptersSafety standards

CB(IEC60601-1/IEC60601-1-11/IEC62368-1), TUV-Mark(EN60601-1/EN60601-1-11 ), UL(ANSI/AAMI ES60601-1/CAN/CSA C22.2 NO.60601-1/UL60950-1/UL62368-1/CAN/CSA C22.2 NO.60950-1), CE(EN62368-1)

      
            

model

Voltage
(volt)

Current
(ampere)

electric power
(W)

Ripple
(maximum)
mVpk-pk

Voltage tolerance

Linear and Negative Adjustment Rates

Average efficiency

Start delay time

UES310D-XXXYYYSPAZ

12.0

19.50

234.0

200

± 5%

 

92.0%

≤3s

15.0

18.00

234.0

200

± 5%

 

92.0%

≤3s

19.0

15.00

285.0

240

± 5%

 

92.0%

≤3s

20.0

14.50

290.0

240

± 5%

Linear adjustment rate:±1%

92.0%

≤3s

24.0

12.50

300.0

300

± 5%

Load:±5%

93.0%

≤3s

48.0

6.46

310.0

400

± 5%

 

93.0%

≤3s

54.0

5.74

310.0

500

± 5%

 

93.0%

≤3s

Input voltage90-264VACWorking temp.0〜40 ℃        
Input Current5.0AStorage temperature-20~60℃       
Output voltage range± 5%Working humidity10~90%RH, no condensation       
Storage humidity5~90% RHmean time between failures > 100,000 hours Telcordia_SR-332, at 25°C ambient temperature.        
weightAbout 800gMechanical specifications73.5 (L)  42.5(W)  30.7(Hmm       
Application placesSuitable for medical and ITE equipment power adaptersSafety standards

CB(IEC60601-1/IEC60601-1-11/IEC62368-1), TUV-Mark(EN60601-1/EN60601-1-11 ), UL(ANSI/AAMI ES60601-1/CAN/CSA C22.2 NO.60601-1/UL60950-1/UL62368-1/CAN/CSA C22.2 NO.60950-1), CE(EN62368-1)

      
            

model

Voltage
(volt)

Current
(ampere)

electric power
(W)

Ripple
(maximum)
mVpk-pk

Voltage tolerance

Linear and Negative Adjustment Rates

Average efficiency

Start delay time

UES310D-XXXYYYSPAZ

12.0

19.50

234.0

200

± 5%

 

92.0%

≤3s

15.0

18.00

234.0

200

± 5%

 

92.0%

≤3s

19.0

15.00

285.0

240

± 5%

 

92.0%

≤3s

20.0

14.50

290.0

240

± 5%

Linear adjustment rate:±1%

92.0%

≤3s

24.0

12.50

300.0

300

± 5%

Load:±5%

93.0%

≤3s

48.0

6.46

310.0

400

± 5%

 

93.0%

≤3s

54.0

5.74

310.0

500

± 5%

 

93.0%

≤3s

Product information

Related Products
model
Output Voltage
Output Current
power
Application industry
Data sheet
Compared
UES300AZ-SPA 19V/20V/24V/48V/54V was added to 0.01-15.7A 300W I.T.E.
UES310D-SPAZ 12V-54V 0.01-19.50 310W Medical/I.T.E.